LNOTE

LNOTE
Name
Code Name
DU Insufficient sample quantity for matrix spike/dup matrix spike LNOTE
DV Corr. coef. for Method of Standard Addition is less than 0.995 LNOTE
DW Sample result is less than reported value LNOTE
DX Value < lowest standard (MQL), but > than MDL LNOTE
DY Compound identified is an analysis at a secondary dilution factor LNOTE
DZ The positive result is an atypical pattern for diesel analysis LNOTE
E EPA Flag - Analyte exceeded the concentration range of the GC/MS LNOTE
EA The positive result is an atypical pattern for gasoline analysis LNOTE
EB Value is estimated LNOTE
EC Value determined by the Method of Standard Addition LNOTE
ED Concentration of total analyte ND; therefore this analyte is ND LNOTE
EF Compound quantitated at a different dilution LNOTE
EG Compound quantitated at a 2x dilution factor LNOTE
EH Compound quantitated at a 5x dilution factor LNOTE
EI Compound quantitated at a 10x dilution factor LNOTE
EJ Compound quantitated at a 20x dilution factor LNOTE
EK Compound quantitated at a 50x dilution factor LNOTE
EL Compound quantitated at a 100x dilution factor LNOTE
EM Compound quantitated at a 200x dilution factor LNOTE
EN Compound quantitated at a 500x dilution factor LNOTE
EO Compound quantitated at a 1000x dilution factor LNOTE
EP Compound quantitated at > 1000x dilution factor LNOTE
EQ Holding time is 28 days by anal. NO3-NO2 on pres. split & sub NO2 LNOTE
ER Thiocyanate not analyzed separately, free CN+SCN below RL free C LNOTE
ES Value undetected at the MDL LNOTE
ET Sample was extracted past end of recommended max. holding time LNOTE
EU LCS is outside of acceptance limits. MS/MDS are acceptable, no correction LNOTE
EV Modified method; see narrative LNOTE
EW Surr. recovery outside of acceptance limits, spike recov. accept. LNOTE
EX Matrix spike diluted to not detectable during analysis LNOTE
EY Result exceeds normal dynamic range; reported as a min. est. LNOTE
EZ Peak coelution, sample concentration may be artificially high LNOTE
FA Peak coelution, sample concentration may be artificially low LNOTE
FB QC ion deviation > 20% relative to labeled std. used for quant. LNOTE
FC Quantitative interference LNOTE
FD Interferences affect 10% or more of the total PCDD/PCDF peak area LNOTE
FE Carbon-labeled int. std. with s/n ratio of < 10:1 LNOTE
FF S/N ratio 10:1 for the int. std. and assoc. analytes LNOTE
FG Results rejected due to int. std. rec. outside QC limits LNOTE
FH Results are valid even though int. std. rec. outside QC limits LNOTE
FI Analyte in field sample and associated blank LNOTE
FJ Total DPE contribution to the total PCDF value > 10% LNOTE
FK EMPC values could be overestimated due to coelution LNOTE
FL Int. std. ion-abundance ratio measured out of accept. range LNOTE
FM Poorly resolved GC peak(s) LNOTE
FN Large closely eluting peak cannot be separated from result LNOTE
FO Est. maximum possible concentration (EMPC) LNOTE
FP Extractable hydrocarbons LNOTE
FQ Gasoline range organics calculated as JP-4 jet fuel LNOTE
FR Diesel range organics, calculated as JP-4 jet fuel LNOTE
FS Diesel range organics, as diesel LNOTE
FT Methanol extraction LNOTE
FU Heated purge LNOTE
FV Purge volume 5 mL LNOTE
FW Purge volume 25 mL LNOTE
FX Analyte present in the instrument blank LNOTE
FY Chromatogram not typical of calibration standard LNOTE
FZ Identified by MS only; based on 1:1 response w/ internal stnd. LNOTE
GA Components not separable by this method; quantified together LNOTE
GB Matrix spike recovery not within control limits LNOTE
GC Second column confirmation performed LNOTE
GD Secondary column result LNOTE
GE Post-digestion spike out of control limits LNOTE
GF Sample absorbance is less than 50% of spike absorbance LNOTE
GG Analytical results not reliable for this common lab contaminant LNOTE
GH Reporting limit must be considered an approximation LNOTE
GI Analyte confirmed by GC/MS LNOTE
GJ Result determined using SIM mode LNOTE
GK Quantitated as an isomer pair LNOTE
GL 2,3,7,8-TCDF results have been confirmed on a DB-225 column LNOTE
GM Result calc. by avg. response factor from init. calib. curve LNOTE
GN Surrogate recovery is outside of control limits LNOTE
GO Result est. below lower calib. limit but above target limit LNOTE
GP Duplicate analysis not within control limits LNOTE
GQ Post-digest. spike recov. 40% - 85% due to matrix interference LNOTE
GR Internal standard recovery is outside method recovery limit LNOTE
GS Reporting limit(s) raised: high level target analyte in sample LNOTE
GT Post-digestion spike recovery <40% due to matrix interference LNOTE
GU >10% breakthrough in second carbon col. due to matrix effects LNOTE
GV Result detected is below the lowest standard and above zero LNOTE
GW Post-digestion spike recovery between 115%-150% LNOTE
GX All reporting limits raised due to high conc. dissolved solids LNOTE
GY Analyte assoc. with sample processing & analysis in lab environ. LNOTE
GZ Secondary ion used for quantitation LNOTE
H A holding time violation has occurred. LNOTE
HA Quantit. of unknown hydrocarbon(s) in sample based on diesel LNOTE
HB Quantit. of unknown hydrocarbon(s) in sample based on gasoline LNOTE
HC Analyte is this compound or an isomer LNOTE
HD Chromat. profile inconsistent with pattern(s) of ref. fuel stnds. LNOTE
HE Extract was run through separation method; see case narrative LNOTE
HF SW3630B clean-up method performed LNOTE
HG Result exceeds calib. range, but not lab's; considered valid LNOTE
HH Result exceeds linear range; concentration may be understated LNOTE
HI SW3611B clean-up method performed LNOTE
HJ Blank spike recovery not within control limits LNOTE
HK Calcium carbonate is greater than 5% of the sample weight LNOTE
HL Analyte recovery above established limit LNOTE
HM Analyte recovery below established limit LNOTE
HN Low concentration matrix spike recovery out of limits LNOTE
HO High concentration matrix spike recovery out of limits LNOTE