LNOTE
Name
Valid Value List
LNOTE Sort ascending NAME
FJ Total DPE contribution to the total PCDF value > 10%
FI Analyte in field sample and associated blank
FH Results are valid even though int. std. rec. outside QC limits
FG Results rejected due to int. std. rec. outside QC limits
FF S/N ratio 10:1 for the int. std. and assoc. analytes
FE Carbon-labeled int. std. with s/n ratio of < 10:1
FD Interferences affect 10% or more of the total PCDD/PCDF peak area
FC Quantitative interference
FB QC ion deviation > 20% relative to labeled std. used for quant.
FA Peak coelution, sample concentration may be artificially low
EZ Peak coelution, sample concentration may be artificially high
EY Result exceeds normal dynamic range; reported as a min. est.
EX Matrix spike diluted to not detectable during analysis
EW Surr. recovery outside of acceptance limits, spike recov. accept.
EV Modified method; see narrative
EU LCS is outside of acceptance limits. MS/DMS are accept., no corr.
ET Sample was extracted past end of recommended max. holding time
ES Value undetected at the MDL
ER Thiocyanate not analyzed separately, free CN+SCN below RL free C
EQ Holding time is 28 days by anal. NO3-NO2 on pres. split & sub NO2
EP Compound quantitated at > 1000x dilution factor
EO Compound quantitated at a 1000x dilution factor
EN Compound quantitated at a 500x dilution factor
EM Compound quantitated at a 200x dilution factor
EL Compound quantitated at a 100x dilution factor
EK Compound quantitated at a 50x dilution factor
EJ Compound quantitated at a 20x dilution factor
EI Compound quantitated at a 10x dilution factor
EH Compound quantitated at a 5x dilution factor
EG Compound quantitated at a 2x dilution factor
EF Compound quantitated at a different dilution
ED Concentration of total analyte ND; therefore this analyte is ND
EC Value determined by the Method of Standard Addition
EB Value is estimated
EA The positive result is an atypical pattern for gasoline analysis
E EPA Flag - Analyte exceeded the concentration range of the GC/MS
DZ The positive result is an atypical pattern for diesel analysis
DY Compound identified is an analysis at a secondary dilution factor
DX Value < lowest standard (MQL), but > than MDL
DW Sample result is less than reported value
DV Corr. coef. for Method of Standard Addition is less than 0.995
DU Insufficient sample quantity for matrix spike/dup matrix spike
DT Method detection limit elevated
DS Batch Quality Assurance data from another project
DR Spiked amount less than five times background level
DQ QC results not applicable because of sample matrix
DP Detected at or above the MRL; refer to reports
DO Coelution
DNQ Detected, Not Quantified
DN 2-Sigma Error
DM Multi-component hydrocarbon mixture
DL Quantified using 30-weight motor oil standard
DK TIC identification based on RT and mass fragmentation pattern
DJ TIC identification based on mass fragmentation pattern only
DI The estimated TIC conc. is < the standard method reporting limit
DH Reporting limits elevated due to insufficient sample quantity
DG Reporting limits elevated due to sample dilution
DF Reporting limits elevated due to matrix interferences
DE Toxic interferences suspected
DD A precipitate formed in the extract during sample preparation
DC Sample produced an emulsion during preparation
DB QA results outside of acceptance limits due to matrix effects
DA Non-fuel components excluded from the results calculation
D EPA Flag - Analytes analyzed at a secondary dilution
CZ Homogeneity could not be readily achieved using routine methods
CY Analysis was performed on a deionized water extract
CX Analysis was performed on a 0.01N HCl extract
CW Carbonate alkalinity is zero because the sample's pH is < 8.3
CV Analysis was performed on a 2M KCl extract
CU Surrogate concentration diluted to not detectable during analysis
CT QC criteria not met due to high level of analyte concentration
CS QC criteria not met due to analyte concentration near RDL
CRF Calibration Response Factor
CR QC criteria not met, sample re-analyzed with similar results
CQA Concentration not reported for QA sample only Percent Recovery or RPD reported.
CQ Analyte conc. greater that 10 times the blank conc.
CP Analyte conc. detectable, but less than 10 times blank conc.
CO Hydrocarbon response in gasoline range but does not resemble gas
CN Hydrocarbon response in diesel range but does not resemble diesel
CM Reporting limits elevated due to low percent solids
CL Initial analysis within holding time but required dilution
CK Initial analysis within holding time but failed QA/QC criteria
CJ Analyte concentration is in excess of the instrument calibration
CI See narrative
CH Sample releases strong petroleum odor
CG Sample releases strong solvent odor
CF Sample releases strong sulfur odor
CE Sample not homogeneous
CD Sample contains multiple phases
CC Sample contains free product
CB Sample contains floculant material
CAL Calculated Value
CA Sample contains white precipitate
C EPA Flag - Pesticide result confirmed using GC/MS
BZ Sample preserved improperly
BY Sample received at improper temperature
BX Sample stored at improper temperature
BW Sample extract analyzed after holding time expired
BV Sample received after holding time expired
BU Sample analyzed after holding time expired

If you notice an error or omission in the valid value list or anywhere on this site, please contact us at help@edfhelpdesk.com - Thank you!